FederalMD
Aug 14, 2026
Closes
Aug 24, 2026
High-Temperature Intrinsic Viscosity Analyzer RFP
High-temperature intrinsic viscosity analyzer. Hardware: Dual capillary/differential viscometer approach — the instrument shall use a dual-capillary measurement of specific viscosity to determine intrinsic viscosity of the polymer — the dimensions (length and radius) of the capillaries and the tolerances of their specification must be provided to NIST to ensure uncertainty propagation can be carried out accurately for the calculation of absolute viscosity by Poiseuille’s Law. Autosampler — instrument shall be able to move hot solutions from vials to the instrument without exposing user to hot solvent. Autosampler shall accommodate at least 20 sample vials. Dissolution method must limit shear degradation to polymers — vendor shall demonstrate limiting sample degradation with dissolution method (either stirring, vortex, or shaking of solvent). Degasser — system shall have integrated degasser to remove air bubbles from solvent. See attached file.