DRAFT Biological Containment Isolation System (BCIS)

Description: Force Protection: Negative pressure differential of 0.01 In.w.g; 12 Air Changes per Hour (ACH); 99.97% at 0.3-micron particle diameter. System Interruption: Visual alarm when pressure is outside the required negative range. Patient Access: Anteroom allows personnel and litter-bound patient ingress/egress while maintaining negative pressure. See attached file.