Procurement of Benchtop Environmental Temp/humidity Chamber RFP
The contractor shall provide an environmental chamber that uniformly conditions samples across a variety of temperature and humidity settings. The chamber must meet the following specifications: Temperature range: at least -20°C to 180°C; Humidity range: 10% to 95% relative humidity (RH); Capacity: 4 CF; Power: Standard power cord and plug (no hardwiring); Water supply: Easily accessible water supply connection; Humidity sensor: Solid-state humidity sensor; Leak Detection: Must have a water leak containment and detection system; Viewing window: For visual monitoring of samples; Interior shelf: At least one internal shelf; Mobility: Mobile cart with space-saving footprint (lab has no bench space and limited floor space for the chamber). See outside link.