Summary
Spectrometer Console: all electronics except NMR probes: Transmitter: A minimum of 3 RF transmitter channels; Minimum of 12.5 ns timing resolution on each channel; Synchronicity on all channels within 12.5 ns; Frequency range for each channel of 15 to 600 MHz (41K to 1H); 1 GB pulse sequencer waveform memory with specific minimum capabiliites of amplitude, phase, and frequency switching--see Table 1; 4 trigger inputs with 12.5 ns resolution;
2 high frequency high power RF amplifiers for 1H and 19F observe and decoupling at 600 and 564 MHz respectively with minimum RF peak power of 500 W; 2 low frequency high power RF amplifiers for X observe and decoupling at 15 to 600 MHz with minimum RF peak power of 500 W. See attached file.